AFOS - Antarctic Fibre Optic Spectrometer
NIGEL - Dome A Fibre Optic Spectrometer
NIGEL AT DOME A


FIBRES:
Types:
|
|
High OH - UV Fibre
|
Low OH - IR Fibre
|
|
| Replacement Set | Ceramoptec | ||
| Currently Installed | Polymicro |
Performance:
|
Product |
Core
(um) ±2% |
Clad
(um) ±2% |
Jacket
(um) ±5% |
Operating
Wavelength (nm) |
|
UV 100/140P
|
100
|
140
|
165
|
180to1100
|
|
WF 100/140P
|
100
|
140
|
165
|
350to2400
|
|
TYPE
|
Attenuation db/50 m
|
|||||||||||||||||
|
Wavelength (nm)
|
280 | 300 | 350 | 400 | 450 | 500 | 550 | 600 | 650 | 700 | 750 | 800 | 850 | 900 | 950 | 1000 | 1050 | 1100 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Ceramoptec OPTRAN UV |
5.25
|
3.75
|
2
|
1.25
|
1.0
|
0.5
|
0.5
|
|
|
|
|
|
|
|
|
|
|
|
| Polymicro FV |
10.5
|
5
|
4
|
1.5
|
1.5
|
0.5
|
0.5
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Ceramoptec OPTRAN WF |
|
|
|
|
|
|
|
0.5
|
0.30
|
0.17
|
0.16
|
0.15
|
0.14
|
0.13
|
0.12
|
0.11
|
0.10
|
0.10
|
| Polymicro FI |
|
|
|
|
|
|
|
0.6
|
0.6
|
0.3
|
0.2
|
0.15
|
0.15
|
0.15
|
0.13
|
0.10
|
0.10
|
0.10
|
CABLES:
CLOSURES & ACCESSORIES
Spectrometer/Slit Module
Injection/Fibre Modules
Drawings & Photo
SPECTROMETER
AFOS AT SOUTH POLE
Injection Module and Cable Splints as installed February 2000

Fibre Optic Cables attached to Injection Module as installed February 2000